Saturday, August 4, 2012

US EPA Environmental Technology Verification Program Approves Use of the Advanced Photonix T-Ray 4000


[logo] US EPAAdvanced Photonix, Inc. Logo


My Note: Thanks to poster bucktailjig on the IV message board for alerting me to this publication. I anticipate API will now move strongly into the use of THz for NDI in the paper industry, matching it's moves into the plastic and nutraceutical markets.

The U.S. Environmental Protection Agency (EPA) has established the Environmental Technology Verification (ETV) Program to facilitate the deployment of innovative or improved environmental technologies through performance verification and dissemination of information. The goal of the ETV Program is to further environmental protection by accelerating the acceptance and use of improved and cost-effective technologies. ETV seeks to achieve this goal by providing high-quality, peer-reviewed data on technology performance to those involved in the design, distribution, financing, permitting, purchase, and use of environmental technologies. Information and ETV documents are available at www.epa.gov/etv.
ETV works in partnership with recognized standards and testing organizations, with stakeholder groups (consisting of buyers, vendor organizations, and permitters), and with individual technology developers. The program evaluates the performance of innovative technologies by developing test plans that are responsive to the needs of stakeholders, conducting field and laboratory tests (as appropriate), collecting and analyzing data, and preparing peer-reviewed reports. All evaluations are conducted according to rigorous quality assurance (QA) protocols to ensure that data of known and adequate quality are generated and that the results are defensible.
The Advanced Monitoring Systems (AMS) Center, one of six verification centers under ETV, is operated by Battelle in cooperation with EPA’s National Risk Management Research Laboratory. The AMS Center evaluated the performance of an alternative technology for sealed radioactive source nuclear gauges. This verification statement provides a summary of the test results for Picometrix T-Ray 4000® Time-Domain Terahertz (TD-THz) System. 
This verification test evaluated the ability of a technology with a non-radioactive source, the Picometrix T-Ray 4000® Time-Domain Terahertz (TD-THz) System, to provide production-line quality control measurements. Testing was designed to monitor basis weight measurements at a paper production facility. Test results were compared to results from a reference technology, the commonly used Measurex MXOpen system nuclear gauge. The test design consisted of a production process phase and a laboratory phase. Testing was conducted on February 16, 17, and 23, 2011 at a commercial paper company in Appleton, WI.
The performance of the technology was tested using two different thicknesses of paper –mid- and heavy weight. For each paper type, five separate rolls (treated as lots at the site) were tested, resulting in ten sets of synchronized nuclear gauge and THz sensor data collected during the actual production process. Data collected within a single lot run were generated within a single roll of product. The emitter for the nuclear gauge system was “parked” at the edge of the paper and the THz sensor aligned with the nuclear gauge emitter. Both the THz sensor and nuclear gauge remained stationary during testing. The data collected throughout each run was time stamped by each technology: the nuclear gauge to 5 second increments and THz results to 0.01 second increments. The nuclear gauge technology was a Measurex MXOpen system that existed at the facility, and was operated by qualified facility staff according to standard operating procedures for that facility. The Picometrix THz technology was operated by the vendor.
A paper sample was collected from the end of each roll used in the production line testing phase. The 10 samples collected from these runs were used for testing during the laboratory phase. Laboratory testing was performed in a Technical Association of the Pulp and Paper Industry (TAPPI) room with tightly controlled temperature and humidity. The paper sample basis weight from each lot was measured using standard TAPPI gravimetric protocols, cut into small squares, stacked, and the time of flight (ToF) measured for each stack using the THz technology. A qualified technician conducted all reference measurements for this verification test according to procedures described in the facility’s Standard Test Method 10001.00 Basis Weight – Laboratory Determination of Coated and Uncoated Paper as modified in the quality assurance project plan (QAPP) for this test.
The Picometrix TD-THz System was verified by evaluating the following parameters:
  • Accuracy – Accuracy was assessed by evaluating the offline basis weight determined by the Picometrix THz sensor on cut, stacked, tear-off paper samples against basis weight measured using the standard laboratory method. The accuracy of the results was assessed by calculating the percent error.
  • Precision – Precision was assessed by triplicate THz sensor measurements of one stack of the heavy paper used for laboratory basis weight determination. Precision was calculated as the standard deviation of repeated measurements.
  • Comparability – Comparability was assessed by evaluating the online results from the nuclear gauge and THz sensor directly against each other on a roll by roll basis. Comparability between the technologies was assessed by calculating the percent difference. A paired t-test was used to determine if the percent difference between technology readings for a given roll of paper was significantly different from zero.
  • Operational Factors – Operational factors such as maintenance needs, power needs, calibration frequency, data output, consumables used, ease of use, repair requirements, training and certification requirements, safety requirements and image throughput were also evaluated.
    QA oversight of verification testing was provided by Battelle and EPA. Battelle QA staff conducted technical systems audits of the testing, and Battelle QA staff conducted a data quality audit of at least 25% of the test data. This verification statement, the full report on which it is based, and the QAPP for this verification test are available at www.epa.gov/etv/centers/center1.html

    Operational Factors. Picometrix states the THz sensor is approximately 50 times smaller and lighter than the nuclear gauge enclosure. The Picometrix LLC THz sensor operates on the principal of EM reflection and refraction which makes system setup, use, and maintenance easier than for the nuclear gauge. Also, the THz system does not rely on any radiological sources and thus does not present any special safety concerns or any special procurement, use, or disposal concerns (e.g., extra disposal costs or potential exposure hazards). The laser must be replaced every 5-10 years; the laser window must be cleaned weekly, but no factory or operator actions are required after the initial vendor setup. The instrument is automatically calibrated during each scan; a NIST standard can be placed in-line monthly at the user’s discretion. The THz system has diagnostics to indicate to the user if and when recalibration is required. The instrument should be adequately grounded when installed, as some occasional interruptions in operation were observed due to static electricity. The instrument requires 90-246 volt power supply and uses < 100 watts of power. Data were output digitally in real-time. The unit cost is $220,000. 



Monday, May 9, 2011

Quality Assurance Project Plan For Verification of Picometrix, LLC T-Ray 4000® Time-Domain Terahertz System


MY NOTE: THIS PDF LINK WILL BE OF GREAT INTEREST TO THE API CONTINGENT, WHO READ THIS BOARD. 

http://www.epa.gov/nrmrl/pubs/600r11037/600r11037.pdf

The following is the introduction from the pdf:


The verification test will be conducted under the U.S. Environmental Protection Agency (EPA) Environmental Technology Verification (ETV) Program. It will be performed by Battelle, which is managing the ETV Advanced Monitoring Systems (AMS) Center through a cooperative agreement with EPA. The scope of the AMS Center covers verification of monitoring technologies for contaminants and natural species in air, water, and soil.
The day to day operations of this verification test will be coordinated and supervised by Battelle, with the participation of the vendor (Picometrix) who will be having the performance of their technology (Picometrix, LLC T-Ray 4000® Time-Domain Terahertz System) verified. The Picometrix technology offers an alternative to using sealed radioactive source nuclear gauges. Testing will be conducted at Appleton Paper in Appleton, Wisconsin. The vendor will provide and operate their technology.

Here is more:

Radioisotopes, such as sealed sources of Strontium-90, Krypton-85 or Promethium-147, can be found in industrial devices used to measure the mass per unit area (basis weight) of various manufactured products such as sheet metal, textiles, paper, photographic film, and other pressed or sheet (flat) material. Devices widely used in industrial and commercial applications are often small in size, and could potentially be lost, stolen, abandoned, or improperly disposed. In some instances sealed radioactive sources can be replaced by an alternative non-radioactive source of energy to accomplish the same function.
Currently, nuclear gauges are being used to measure or control material density, flow, level, thickness or basis weight. The gauges contain sealed sources that radiate through the substance being measured to a readout or controlling device. The radiation that is emitted from the gauge is attenuated by the matter between the radioisotope and the detector. The radiation reduction can be correlated to the quantity of matter between the source and the detector. The
nuclear radiation is attenuated by the number of atoms between the source and the detector. The physical measurement is the magnitude of radiation that passes through the material. The actual amplitude of the detected signal must be compensated for a number of factors (e.g. moisture, air in the intervening space, geometry of the source and detector); however these factors are well understood and implemented in commercial nuclear gauges.
Nuclear gauges have been widely used (since the 1940’s) for the measurement of basis weight and quantities that can be derived from basis weight (e.g. thickness, coat weight) because of their ability to penetrate most materials, their long life (depending on isotope) and their low cost. At this point in time they are a well established, low risk gauge that is well known in the industry. However, the radioactive sources in these gauges can present a safety risk. Minimizing the number of radioactive sources in the public domain will decrease the opportunity for individuals to use the radioisotopes for unintended, harmful (i.e. terrorist) purposes. In an effort to do so, the EPA's Office of Radiation and Indoor Air in the Office of Air and Radiation (OAR), established the EPA's Alternative Technologies Initiative (ATI) (http://www.epa.gov/radiation/source-reduction-management/alt-technologies.html). Part of the EPA-ATI is fostering the acceptance and voluntary market adoption of non-radioactive technologies; i.e., alternative technologies to those that currently use sealed sources. The EPA-ATI is focusing primarily on alternative technologies for devices with Category 3 and 4 radioactive sources as classified by the International Atomic Energy Agency (IAEA) . Commercial-ready or available alternatives to nuclear gauges (such as infrared, laser, and other technologies) are being considered. As with any new technology, the likelihood of acceptance can be significantly increased by independent evaluation and verification of a technology's capabilities.
The purpose of this QAPP is to specify procedures for a verification test applicable to commercially available alternatives to nuclear gauges that can replace technologies using sealed radioactive sources for the measurement of basis weight, the measurement of mass per unit area of a material. The purpose of the verification test is to evaluate the performance of the Picometrix T-Ray 4000® Time-Domain Terahertz (THz) System in an actual production environment for the measurement of basis weight.